Potentiostat

  • Compliance voltage: ±12 V
  • Applied potential: ±10 V FSR
  • Resolution: 32-bit (4.7 nV)
  • Measured potential: ±10 V FSR
  • Resolution: 24-bit (1.2 μV)
  • Current ranges: 10-decades 1 nA to 1 A
  • Maximum current: 500 mA
  • Current resolution: 23.8 fA
  • Accuracy: <0.5%
  • Floating capability Standard
  • Cell connections 2, 3 or 4
  • Maximum ADC sample rate: 4 MHz
  • Maximum ADC resolution: 24-bit
  • Minimum pulse duration: 100 μs
  • Scan rate: 1 μV/s to 200 V/s

ic-SECM module

  • Piezo crystal extension 100 µm
  • Vibration frequency 80 – 600 Hz
  • Vibration control 20 nm – 2 µm peak to peak
  • Minimum increment 1 nm
  • Z control resolution 0.09 nm (piezo)
  • Topography resolution (recorded) 20 nm

Positioning system specifications

Stepper Motors

  • Scan Range (x,y,z): 110 mm x 110 mm x 110 mm
  • Minimal step size on all axes: 20 nm
  • Closed loop positioning linear zero hysteresis encoder with direct real-time readout of displacement in x, y and z
  • Linear position encoder resolution: 20 nm.
  • Max. scan speed: 10 mm/s
  • Measurement resolution: 32-bit decoder @ up to 40 MHz

Piezoelectric element (for z axis only)

  • Vibration range 20 nm – 2 µm peak to peak with 1 nm increments
  • Min. vibration resolution: 0.12 nm calculated (16-bit DAC on 4 µm)
  • Piezo crystal extension: 100 µm
  • Positioning resolution: 0.09 nm calculated (20-bit DAC on 100 µm)