Specifications Potentiostat Compliance voltage: ±12 V Applied potential: ±10 V FSR Resolution: 32-bit (4.7 nV) Measured potential: ±10 V FSR Resolution: 24-bit (1.2 μV) Current ranges: 10-decades 1 nA to 1 A Maximum current: 500 mA Current resolution: 23.8 fA Accuracy: >0.5% Floating capability Standard Cell connections 2, 3 or 4 Maximum ADC sample rate: 4 MHz Maximum ADC resolution: 24-bit Minimum pulse duration: 100 μs Scan rate: 1 μV/s to 200 V/s EIS Capability Frequency range: 1 µHz to 1 MHz Analyser accuracy: 0.1%, 0.1° Frequency resolution 66 nHz Mode Single sine, Multisine, FFT analysis Positioning system specifications Stepper Motors Scan Range (x,y,z): 110 mm x 110 mm x 110 mm Minimal step size on all axes: 20 nm Closed loop positioning linear zero hysteresis encoder with direct real-time readout of displacement in x, y and z Linear position encoder resolution: 20 nm. Max. scan speed: 10 mm/s Measurement resolution: 32-bit decoder @ up to 40 MHz Piezoelectric element (for z axis only) Vibration range 20 nm – 2 µm peak to peak with 1 nm increments Min. vibration resolution: 0.12 nm calculated (16-bit DAC on 4 µm) Piezo crystal extension: 100 µm Positioning resolution: 0.09 nm calculated (20-bit DAC on 100 µm) Read more …
Specifications Positioning system • Scan range : X&Y = 200 μm; Z = 100 μm • Practical minimum step size : 50 nm with <10 nm resolution • Max scan speed : 200 μm/s • Max discrete data acquisition for line or area scan : 20 points/s • Micrometer macro positioning : 13 mm range – the smallest graduation is 10 μm • Max area scan point density : >4,000,000 Potentiostats • Current measurement : 2.56 x 1 mA to 100 pA with 0.5% accuracy (2% for 100 pA) and resolution (2 x range @ 16 bit, 7.8 fA @ 100 pA) • Potential : ±2.048 V for applied voltage with 0.5% accuracy and 62 μV resolution, ±2.56 V for measured voltage with 0.5% accuracy and 78 μV resolution • Electrometer : <10 pA input bias current and 1011 Ω II 5 pF input impedance (R/C) • Time base : typically 100 μs • Sampling rate : 1 mHz to 10 kHz • Maximum data samples : >100,000 (technique dependent) • Electrical Isolation : yes • Power : via USB • PC connection : USB • Cell connections : with 2 mm banana plugs(WE=Red, RE=White, CE=Blue) Probes • Probes options from 25 µm – 1 µm General • Control box dimensions : 90 x 235 x 180 mm (H x W x D) • Scanning stage dimensions : 190 x 225 x 230 mm (H x W x D) • Potentiostat dimensions : 97 x 15 x 54 mm (H x W x D) • Potentiostat weight : 61 g • Total weight : 8 kg • Operating temperature range : 25 °C ±5 °C • Operating humidity range : 75% ±10% • Power : 100 to 240 V, 50/60 Hz, 60 W (Universal external certified power pack) Read more …